Philips recalls CPAP machines
For patients impacted by the Philips recall, the below links will provide more information.
To begin a claim, visit:
https://www.philipssrcupdate.expertinquiry.com/
Philips Respironics has issued a recall on thousands of ventilators and CPAP machines. The recall only affects units sold in the United States.
The units affected include specific Philips Bi-Level Positive Airway Pressure (Bi-Level PAP), Continuous Positive Airway Pressure (CPAP), and mechanical ventilators. The issue stems from potential health risks related to the polyester-based polyurethane (PE-PUR) sound abatement foam component in these devices.
Philips says the foam has a chance to break down and be inhaled by device users, the foam may off-gas certain chemicals. They say the deterioration could intensify by using unapproved cleaning methods and high heat and high humidity environments.
The potential risks of particulate exposure include headache, irritation, inflammation, respiratory issues, and possible toxic and carcinogenic effects. The potential risks of chemical exposure due to off-gassing include headache, irritation, hypersensitivity, nausea/vomiting, and possible toxic and carcinogenic effects.
Philips has received some complaints of possible patient impact due to foam degradation. There have not received any reports of death or patient impact related to chemical emissions.
Customers and patients should take the following action:
Based on the latest analysis of potential health risks and out of an abundance of caution, the recall notification* advises patients and customers to take the following actions:
- For patients using affected BiLevel PAP and CPAP devices: Discontinue use of your device and work with your physician or Durable Medical Equipment (DME) provider to determine the most appropriate options for continued treatment. To continue use of your device due to lack of alternatives, consult with your physician to determine if the benefit of continuing therapy with your device outweighs the risks identified in the recall notification.*
- For patients using affected life-sustaining mechanical ventilator devices: Do not stop or alter your prescribed therapy until you have talked to your physician. Philips recognizes that alternate ventilator options for therapy may not exist or may be severely limited for patients who require a ventilator for life-sustaining therapy, or in cases where therapy disruption is unacceptable. In these situations, and at the discretion of the treating clinical team, the benefit of continued usage of these ventilator devices may outweigh the risks identified in the recall notification.*
Possible health risks
The company continues to monitor reports of potential safety issues as required by medical device regulations and laws in the markets in which it operates. To date, there have been no reports of death as a result of these issues. Philips has received reports of possible patient impact due to foam degradation. The potential risks of particulate exposure include headache, irritation, inflammation, respiratory issues, and possible toxic and carcinogenic effects. The potential risks of chemical exposure due to off-gassing include headache, irritation, hypersensitivity, nausea/vomiting, and possible toxic and carcinogenic effects. Philips has received no reports regarding patient impact related to chemical emissions.
Repair and replacement program
Philips is providing the relevant regulatory agencies with required information related to the launch and implementation of the projected correction. The company will replace the current sound abatement foam with a new material and has already begun the preparations, which include obtaining the relevant regulatory clearances. Philips aims to address all affected devices in scope of this correction as expeditiously as possible.
As part of the program, the first-generation DreamStation product families will be modified with a different sound abatement foam and shipped upon receipt of the required regulatory clearances. Philips’ recently launched next-generation CPAP platform, DreamStation 2, is not affected by the issue. To support the program, Philips is increasing the production of its DreamStation 2 CPAP devices, that are available in the US and selected countries in Europe.
Additional information
For more information on the recall notification,* as well as instructions for customers, users and physicians, affected parties may contact their local Philips representative or visit www.philips.com/SRC-update.